Quant Specifications
Spectroscopic Performance (typical, at 25° Celsius with InAs detector)
Spectral Range 3,700 to 14,885 cm-1 (672 to 2702 nm)
Signal to Noise (60 s, 16 cm-1, at peak response) > 900 000: 1
Frequency Repeatability (@ 7300 cm-1) < 0.006 cm-1
Frequency Accuracy (@ 7300 cm-1) < 0.01 cm-1
Absorbance Reproducibility (toluene) < 0.002 AU
Optical Bench
Beamsplitter Material ZnSe (non-hygroscopic).
Interferometer Mechanism (patent pending) High-throughput double pivot Michelson, fully jacketed.
Source (with electronic stabilization) Quartz Halogen, 10-year expected life.
VCSEL Solid-state Laser - Maintenance Free.
Data Communications
Hardware Port Ethernet, 10/100 Mbps
Instrument Enclosure
Material Rugged Design, all-metal frame with integral handles
Size 435 mm (W) x 280 mm (D) x 372 mm (H), Weight 24 kg
Environmental
Universal Power Supply 120-240 V AC, 50/60 Hz, 65 W
Operating Temperature 10°C to 35°C
Operating Relative Humidity 5% to 80%, non-condensing
Regulatory Certification and Compliance TUV, CE and RoHS
For further information on this product range please contact Mark Whatton, FT-NIR Applications Manager, on 01342 820820 or email enquiries@qclscientific.com.
Quant Main Page Configurations Maintenance InfraQuant 2.0 Software
FT-NIR Applications FT-NIR Technology
Categories
|